Your browser doesn't support javascript.
Show: 20 | 50 | 100
Results 1 - 20 de 101
Filter
2.
PLoS One ; 17(11): e0268863, 2022.
Article in English | MEDLINE | ID: covidwho-2112689

ABSTRACT

BACKGROUND: The reprocessing of daily used medical devices is often inadequate, making them a potential source of infection. In addition, there are usually no consistent and technically standardized procedures available for this purpose. Hence, the aim of this study is to analyze the bacterial contamination and the effectiveness of Ultraviolet light-based (UV light-based) reprocessing of daily used medical devices. MATERIAL AND METHODS: Six different everyday medical devices (20 each; stethoscopes, tourniquets, bandage scissors, reflex hammers, tuning forks, and nystagmus glasses) were tested for bacterial contamination. All medical devices were then exposed to UV-C light for 25 seconds. Medical devices with a smooth surface were pre-cleaned with a water-based wipe. Contact samples were taken before and after reprocessing. RESULTS: Immediately after clinical use, 104 of 120 contact samples showed an average bacterial contamination of 44.8±64.3 colony forming units (CFU) (0-300 CFU), also including potentially pathogenic bacteria. Two further culture media were completely overgrown with potentially pathogenic bacteria. The stethoscopes were found to have the highest average contamination of 90±91.6 CFU. After reprocessing, 118 of 120 samples were sterile, resulting in an average residual contamination of 0.02±0.1 CFU in two samples, whereby only bacteria of the ordinary skin flora were found. CONCLUSION: The present study shows the potentially clinically relevant bacterial contamination of everyday used medical devices. The reprocessing method tested here using UV light appears to be a suitable method for disinfection, especially for objects that up to now have been difficult to disinfect or cannot be disinfected in a standardized manner.


Subject(s)
Equipment Contamination , Ultraviolet Rays , Equipment Contamination/prevention & control , Disinfection/methods , Bacteria , Drug Contamination
4.
PLoS One ; 17(10): e0275488, 2022.
Article in English | MEDLINE | ID: covidwho-2054385

ABSTRACT

Glutaraldehyde, a germicide for reprocessing endoscopes that is important for hygiene in the clinic, might be hazardous to humans. Electrolyzed acid water (EAW) has a broad anti-microbial spectrum and safety profile and might be a glutaraldehyde alternative. We sought to assess EAW disinfection of flexible endoscopes in clinical otorhinolaryngological settings and its in vitro inactivation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and bacteria commonly isolated in otorhinolaryngology. Ninety endoscopes were tested for bacterial contamination before and after endoscope disinfection with EAW. The species and strains of bacteria were studied. The in vitro inactivation of bacteria and SARS-CoV-2 by EAW was investigated to determine the efficacy of endoscope disinfection. More than 20 colony-forming units of bacteria at one or more sampling sites were detected in 75/90 microbiological cultures of samples from clinically used endoscopes (83.3%). The most common genus detected was Staphylococcus followed by Cutibacterium and Corynebacterium at all sites including the ears, noses, and throats. In the in vitro study, more than 107 CFU/mL of all bacterial species examined were reduced to below the detection limit (<10 CFU/mL) within 30 s after contact with EAW. When SARS-CoV-2 was treated with a 99-fold volume of EAW, the initial viral titer (> 105 PFU) was decreased to less than 5 PFU. Effective inactivation of SARS-CoV-2 was also observed with a 19:1 ratio of EAW to the virus. EAW effectively reprocessed flexible endoscopes contributing to infection control in medical institutions in the era of the coronavirus disease 2019 pandemic.


Subject(s)
COVID-19 , Disinfection , Bacteria , COVID-19/prevention & control , Cross-Sectional Studies , Endoscopes/microbiology , Endoscopes, Gastrointestinal/microbiology , Equipment Contamination/prevention & control , Glutaral , Humans , SARS-CoV-2 , Water
5.
Int J Environ Res Public Health ; 19(17)2022 Aug 24.
Article in English | MEDLINE | ID: covidwho-2023688

ABSTRACT

In dental clinics, the infections may be acquired through contaminated devices, air, and water. Aerosolized water may contain bacteria, grown into the biofilm of dental unit waterlines (DUWLs). We evaluated a disinfection method based on water osmosis and chlorination with chlorine dioxide (O-CD), applied to DUWL of five dental clinics. Municipal water was chlorinated with O-CD device before feeding all DUWLs. Samplings were performed on water/air samples in order to research total microbial counts at 22-37 °C, Pseudomonas aeruginosa, Legionella spp., and chlorine values. Water was collected from the taps, spittoons, and air/water syringes. Air was sampled before, during, and after 15 min of aerosolizing procedure. Legionella and P. aeruginosa resulted as absent in all water samples, which presented total microbial counts almost always at 0 CFU/mL. Mean values of total chlorine ranged from 0.18-0.23 mg/L. Air samples resulted as free from Legionella spp. and Pseudomonas aeruginosa. Total microbial counts decreased from the pre-aerosolizing (mean 2.1 × 102 CFU/m3) to the post-aerosolizing samples (mean 1.5 × 10 CFU/m3), while chlorine values increased from 0 to 0.06 mg/L. O-CD resulted as effective against the biofilm formation in DUWLs. The presence of residual activity of chlorine dioxide also allowed the bacteria reduction from air, at least at one meter from the aerosolizing source.


Subject(s)
Disinfection , Legionella , Bacteria , Biofilms , Chlorine/pharmacology , Chlorine Compounds , Colony Count, Microbial , Dental Equipment , Disinfection/methods , Equipment Contamination/prevention & control , Osmosis , Oxides , Pseudomonas aeruginosa , Water , Water Microbiology
6.
Respiration ; 100(12): 1208-1217, 2021.
Article in English | MEDLINE | ID: covidwho-1956956

ABSTRACT

Bronchoscopy is a commonly performed procedure within thoracic and critical care medicine. Modern bronchoscopes are technologically advanced tools made of fragile electronic components. Their design is catered to allow maximum maneuverability within the semi-rigid tracheobronchial tree. Effective cleaning and reprocessing of these tools can be a challenge. Although highly functional, the design poses several challenges when it comes to reprocessing. It is a very important step, and lapses in the procedure have been tied to nosocomial infections. The process lacks universal standardization; several organizations have developed their own recommendations. Data have shown that key stakeholders are not fully versed in the essentials of endoscope reprocessing. A significant knowledge gap exists between those performing bronchoscopy and those who are stewards of effective endoscope reprocessing. To service as a resource for bronchoscopists, this study summarizes the steps of effective reprocessing, details the important elements within a health-care facility that houses this process, and reviews some of the current data regarding the use of disposable endoscopes.


Subject(s)
Disinfection , Equipment Contamination , Bronchoscopes , Disinfection/methods , Endoscopes , Equipment Contamination/prevention & control , Humans , Reference Standards
8.
ACS Appl Mater Interfaces ; 14(9): 11068-11077, 2022 Mar 09.
Article in English | MEDLINE | ID: covidwho-1713108

ABSTRACT

Amidst the COVID-19 pandemic, it is evident that viral spread is mediated through several different transmission pathways. Reduction of these transmission pathways is urgently needed to control the spread of viruses between infected and susceptible individuals. Herein, we report the use of pathogen-repellent plastic wraps (RepelWrap) with engineered surface structures at multiple length scales (nanoscale to microscale) as a means of reducing the indirect contact transmission of viruses through fomites. To quantify viral repellency, we developed a touch-based viral quantification assay to mimic the interaction of a contaminated human touch with a surface through the modification of traditional viral quantification methods (viral plaque and TCID50 assays). These studies demonstrate that RepelWrap reduced contamination with an enveloped DNA virus as well as the human coronavirus 229E (HuCoV-229E) by more than 4 log 10 (>99.99%) compared to a standard commercially available polyethylene plastic wrap. In addition, RepelWrap maintained its repellent properties after repeated 300 touches and did not show an accumulation in viral titer after multiple contacts with contaminated surfaces, while increases were seen on other commonly used surfaces. These findings show the potential use of repellent surfaces in reducing viral contamination on surfaces, which could, in turn, reduce the surface-based spread and transmission.


Subject(s)
COVID-19/prevention & control , Coronavirus 229E, Human/growth & development , Equipment Contamination/prevention & control , Infection Control/instrumentation , Plastics/chemistry , COVID-19/transmission , COVID-19/virology , Humans , Infection Control/methods , SARS-CoV-2/growth & development , Surface Properties
9.
Int J Mol Sci ; 23(3)2022 Feb 08.
Article in English | MEDLINE | ID: covidwho-1686816

ABSTRACT

This Special Issue of the International Journal of Molecular Sciences, entitled "Antimicrobial Materials with Medical Applications", covers a selection of recent research and review articles in the field of antimicrobial materials, as well as their medical applications [...].


Subject(s)
Anti-Infective Agents/pharmacology , Disinfectants/pharmacology , Equipment Contamination/prevention & control , Drug Development , Drug Resistance, Microbial , Humans , Product Packaging
10.
Sci Rep ; 11(1): 22868, 2021 11 24.
Article in English | MEDLINE | ID: covidwho-1532097

ABSTRACT

Transfer of SARS-CoV-2 from solids to fingers is one step in infection via contaminated solids, and the possibility of infection from this route has driven calls for increased frequency of handwashing during the COVID-19 pandemic. To analyze this route of infection, we measured the percentage of SARS-CoV-2 that was transferred from a solid to an artificial finger. A droplet of SARS-CoV-2 suspension (1 µL) was placed on a solid, and then artificial skin was briefly pressed against the solid with a light force (3 N). Transfer from a variety of solids was detected, and transfer from the non-porous solids, glass, stainless steel, and Teflon, was substantial when the droplet was still wet. The viral titer for the finger was 13-16% or 0.8-0.9 log less than for the input droplet. Transfer still occurred after the droplet evaporated, but was smaller, 3-9%. We found a lower level of transfer from porous solids but did not find a significant effect of solid wettability for non-porous solids.


Subject(s)
COVID-19/transmission , Disease Transmission, Infectious/prevention & control , SARS-CoV-2/metabolism , COVID-19/metabolism , Equipment Contamination/prevention & control , Equipment Contamination/statistics & numerical data , Humans , SARS-CoV-2/pathogenicity , Skin/virology , Viral Load
11.
J Hosp Infect ; 113: 104-114, 2021 Jul.
Article in English | MEDLINE | ID: covidwho-1531580

ABSTRACT

Healthcare-associated infections (HAIs) are the most common adverse outcomes due to delivery of medical care. HAIs increase morbidity and mortality, prolong hospital stay, and are associated with additional healthcare costs. Contaminated surfaces, particularly those that are touched frequently, act as reservoirs for pathogens and contribute towards pathogen transmission. Therefore, healthcare hygiene requires a comprehensive approach whereby different strategies may be implemented together, next to targeted, risk-based approaches, in order to reduce the risk of HAIs for patients. This approach includes hand hygiene in conjunction with environmental cleaning and disinfection of surfaces and clinical equipment. This review focuses on routine environmental cleaning and disinfection including areas with a moderate risk of contamination, such as general wards. As scientific evidence has not yet resulted in universally accepted guidelines nor led to universally accepted practical recommendations pertaining to surface cleaning and disinfection, this review provides expert guidance for healthcare workers in their daily practice. It also covers outbreak situations and suggests practical guidance for clinically relevant pathogens. Key elements of environmental cleaning and disinfection, including a fundamental clinical risk assessment, choice of appropriate disinfectants and cleaning equipment, definitions for standardized cleaning processes and the relevance of structured training, are reviewed in detail with a focus on practical topics and implementation.


Subject(s)
Cross Infection , Disinfectants , Cross Infection/prevention & control , Delivery of Health Care , Disinfection , Equipment Contamination/prevention & control , Humans , Hygiene
12.
Infect Dis Clin North Am ; 35(3): 631-666, 2021 09.
Article in English | MEDLINE | ID: covidwho-1340080

ABSTRACT

Outbreaks and pseudo-outbreaks in health care settings are complex and should be evaluated systematically using epidemiologic and molecular tools. Outbreaks result from failures of infection prevention practices, inadequate staffing, and undertrained or overcommitted health care personnel. Contaminated hands, equipment, supplies, water, ventilation systems, and environment may also contribute. Neonatal intensive care, endoscopy, oncology, and transplant units are areas at particular risk. Procedures, such as bronchoscopy and endoscopy, are sources of infection when cleaning and disinfection processes are inadequate. New types of equipment can be introduced and lead to contamination or equipment and medications can be contaminated at the manufacturing source.


Subject(s)
Cross Infection/prevention & control , Delivery of Health Care/organization & administration , Disease Outbreaks/prevention & control , Disinfection , Equipment Contamination/prevention & control , Infection Control , Health Facilities , Humans , Infant, Newborn
13.
Infect Dis Clin North Am ; 35(3): 575-607, 2021 09.
Article in English | MEDLINE | ID: covidwho-1340079

ABSTRACT

All invasive procedures involve contact by a medical device or surgical instrument with a patient's sterile tissue or mucous membranes. The level of disinfection is dependent on the intended use of the object: critical, semicritical, or noncritical. New issues and practices can affect the risk of infection associated with devices and surfaces. Endoscopes continue to represent a nosocomial hazard. The contaminated surface environment in hospital rooms is important in the transmission of health care-associated pathogens. Thoroughness of cleaning must be monitored and no-touch room decontamination technology should be. In general, emerging pathogens are susceptible to currently available disinfectants.


Subject(s)
Cross Infection/prevention & control , Delivery of Health Care/organization & administration , Disinfectants , Disinfection/methods , Equipment Contamination/prevention & control , Sterilization/methods , Endoscopes , Equipment and Supplies, Hospital , Hospitals , Humans , Temperature
14.
BMJ Glob Health ; 5(10)2020 10.
Article in English | MEDLINE | ID: covidwho-1388494

ABSTRACT

INTRODUCTION: During pandemics, such as the SARS-CoV-2, filtering facepiece respirators plays an essential role in protecting healthcare personnel. The recycling of respirators is possible in case of critical shortage, but it raises the question of the effectiveness of decontamination as well as the performance of the reused respirators. METHOD: Disposable respirators were subjected to ultraviolet germicidal irradiation (UVGI) treatment at single or successive doses of 60 mJ/cm2 after a short drying cycle (30 min, 70°C). The germicidal efficacy of this treatment was tested by spiking respirators with two staphylococcal bacteriophages (vB_HSa_2002 and P66 phages). The respirator performance was investigated by the following parameters: particle penetration (NaCl aerosol, 10-300 nm), scanning electron microscopy (SEM), Fourier-transform infrared spectroscopy (FTIR), differential scanning calorimetry and mechanical tensile tests. RESULTS: No viable phage particles were recovered from any of the respirators after decontamination (log reduction in virus titre >3), and no reduction in chemical or physical properties (SEM, particle penetrations <5%-6%) were observed. Increasing the UVGI dose 10-fold led to chemical alterations of the respirator filtration media (FTIR) but did not affect the physical properties (particle penetration), which was unaltered even at 3000 mJ/cm2 (50 cycles). When respirators had been used by healthcare workers and undergone decontamination, they had particle penetration significantly greater than never donned respirators. CONCLUSION: This decontamination procedure is an attractive method for respirators in case of shortages during a SARS pandemic. A successful implementation requires a careful design and particle penetration performance control tests over the successive reuse cycles.


Subject(s)
Decontamination/methods , Equipment Contamination/prevention & control , Equipment Reuse , Respiratory Protective Devices , Ultraviolet Rays , Betacoronavirus , COVID-19 , Coronavirus Infections/prevention & control , Equipment Failure Analysis , Humans , Infection Control/methods , Materials Testing , Pandemics/prevention & control , Pneumonia, Viral/prevention & control , SARS-CoV-2
15.
Cytometry A ; 97(9): 882-886, 2020 09.
Article in English | MEDLINE | ID: covidwho-1384154

ABSTRACT

Operating shared resource laboratories (SRLs) in times of pandemic is a challenge for research institutions. In a multiuser, high-turnover working space, the transmission of infectious agents is difficult to control. To address this challenge, imaging core facility managers being members of German BioImaging discussed how shared microscopes could be operated with minimal risk of spreading SARS-CoV-2 between users and staff. Here, we describe the resulting guidelines and explain their rationale, with a focus on separating users in space and time, protective face masks, and keeping surfaces virus-free. These recommendations may prove useful for other types of SRLs. © 2020 The Authors. Cytometry Part A published by Wiley Periodicals LLC. on behalf of International Society for Advancement of Cytometry.


Subject(s)
Betacoronavirus/pathogenicity , Biomedical Research/organization & administration , Coronavirus Infections/prevention & control , Infection Control , Laboratories/organization & administration , Microscopy , Occupational Health , Pandemics/prevention & control , Pneumonia, Viral/prevention & control , COVID-19 , Cooperative Behavior , Coronavirus Infections/transmission , Coronavirus Infections/virology , Decontamination , Equipment Contamination/prevention & control , Germany , Humans , Occupational Exposure/prevention & control , Personal Protective Equipment , Pneumonia, Viral/transmission , Pneumonia, Viral/virology , Protective Factors , Research Personnel/organization & administration , Risk Assessment , Risk Factors , SARS-CoV-2 , Workflow
16.
Antimicrob Resist Infect Control ; 10(1): 109, 2021 07 23.
Article in English | MEDLINE | ID: covidwho-1365395

ABSTRACT

BACKGROUND: Laparoscopy is a minimally-invasive surgical procedure that uses long slender instruments that require much smaller incisions than conventional surgery. This leads to faster recovery times, fewer post-surgical wound infections and shorter hospital stays. For these reasons, laparoscopy could be particularly advantageous to patients in low to middle income countries (LMICs). Unfortunately, sterile processing departments in LMIC hospitals are faced with limited access to equipment and trained staff which poses an obstacle to safe surgical care. The reprocessing of laparoscopic devices requires specialised equipment and training. Therefore, when LMIC hospitals invest in laparoscopy, an update of the standard operating procedure in sterile processing is required. Currently, it is unclear whether LMIC hospitals, that already perform laparoscopy, have managed to introduce updated reprocessing methods that minimally invasive equipment requires. The aim of this study was to identify the laparoscopic sterile reprocessing procedures in rural India and to test the effectiveness of the sterilisation equipment. METHODS: We assessed laparoscopic instrument sterilisation capacity in four rural hospitals in different states in India using a mixed-methods approach. As the main form of data collection, we developed a standardised observational checklist based on reprocessing guidelines from several sources. Steam autoclave performance was measured by monitoring the autoclave cycles in two hospitals. Finally, the findings from the checklist data was supported by an interview survey with surgeons and nurses. RESULTS: The checklist data revealed the reprocessing methods the hospitals used in the reprocessing of laparoscopic instruments. It showed that the standard operating procedures had not been updated since the introduction of laparoscopy and the same reprocessing methods for regular surgical instruments were still applied. The interviews confirmed that staff had not received additional training and that they were unaware of the hazardous effects of reprocessing detergents and disinfectants. CONCLUSION: As laparoscopy is becoming more prevalent in LMICs, updated policy is needed to incorporate minimally invasive instrument reprocessing in medical practitioner and staff training programmes. While reprocessing standards improve, it is essential to develop instruments and reprocessing equipment that is more suitable for resource-constrained rural surgical environments.


Subject(s)
Equipment Contamination/prevention & control , Hospitals, Rural , Laparoscopy , Sterilization/methods , Developing Countries , India , Steam , Sterilization/instrumentation
17.
Antimicrob Resist Infect Control ; 10(1): 120, 2021 08 16.
Article in English | MEDLINE | ID: covidwho-1365394

ABSTRACT

BACKGROUND: The hospital environment has got more attention as evidence as source for bacterial transmission and subsequent hospital-acquired infection increased. Regular cleaning and disinfection have been proposed to lower the risk of infection, in particular for gram-positive bacteria. Auto-disinfecting surfaces would allow to decrease survival of pathogens, while limiting resource to achieve a safe environment in patient rooms. METHODS: A controlled trial to evaluate the antimicrobial effectiveness of a polyvinyl chloride foil containing an integrated silver-based agent (containing silver ions 2%) on high-touch surfaces in patient rooms. RESULTS: The overall log reduction of the mean values was 1.8 log10 CFU, the median 0.5 log10 CFU comparing bioburden of control vs antimicrobial foil (p < 0.01). Important pathogens were significantly less likely recovered from the foil, in particular enterococci. These effects were present even after 6 months of in-use. CONCLUSIONS: A foil containing an integrated silver-based agent applied to high-touch surfaces effectively results in lower recovery of important pathogens from such surfaces over a 6-month study period.


Subject(s)
Cross Infection/prevention & control , Disinfectants/pharmacology , Disinfection/methods , Equipment Contamination/prevention & control , Patients' Rooms , Silver/pharmacology , Fomites/microbiology , Hospitals , Polyvinyl Chloride , Prospective Studies , Switzerland , Touch
18.
BMC Infect Dis ; 21(1): 681, 2021 Jul 13.
Article in English | MEDLINE | ID: covidwho-1317122

ABSTRACT

BACKGROUND: Mobile phones used by healthcare workers (HCWs) are contaminated with bacteria, but the posterior surface of smartphones has rarely been studied. The aim of this study was to compare the prevalence of microbial contamination of touchscreens and posterior surfaces of smartphones owned by HCWs. METHODS: A cross-sectional study of smartphones used by HCWs employed at two intensive care units at a Japanese tertiary care hospital was performed. Bacteria on each surface of the smartphones were isolated separately. The primary outcomes were the prevalence of microbial contamination on each surface of smartphones and associated bacterial species. Fisher's exact test was used to compare dichotomous outcomes. RESULTS: Eighty-four HCWs participated in this study. The touchscreen and posterior surface were contaminated in 27 (32.1%) and 39 (46.4%) smartphones, respectively, indicating that the posterior surface was more frequently contaminated (p = 0.041). Bacillus species and coagulase-negative staphylococci were isolated from each surface of the smartphones. CONCLUSIONS: The posterior surface of a smartphone was more significantly contaminated with bacteria than the touchscreen, regardless of having a cover. Therefore, routine cleaning of the posterior surface of a smartphone is recommended.


Subject(s)
Bacillus/isolation & purification , Equipment Contamination , Health Personnel/statistics & numerical data , Smartphone , Staphylococcus/isolation & purification , Cross Infection/prevention & control , Cross-Sectional Studies , Equipment Contamination/prevention & control , Equipment Contamination/statistics & numerical data , Humans , Infection Control/methods , Intensive Care Units/statistics & numerical data , Japan , Prevalence
19.
J Prev Med Hyg ; 62(1): E104-E109, 2021 Mar.
Article in English | MEDLINE | ID: covidwho-1264705

ABSTRACT

BACKGROUND: Healthcare workers' (HCW) hands and personnel belongings are vehicles of transmission of nosocomial infections. Knowledge, attitude, and practice of hand hygiene have been extensively studied suggesting adequate knowledge but poor compliance. Similar data on aprons, mobile phone and stethoscope disinfection practices are lacking. This becomes an extensively important topic of discussion in current COVID-19 pandemic where inadequacy in hygiene practices is devastating. AIM: To study the knowledge, attitude, and infection prevention practices of HCWs aprons, electronic devices, stethoscopes, and hands. METHODS: A cross sectional questionnaire-based survey was conducted among HCWs of Medicine ward and ICU. RESULTS: Sixty-six HCWs responded to the survey. Awareness that hands, aprons, mobile phones, stethoscopes could cause cross transmission and knowledge of correct practices was present in majority of the respondents. Hand hygiene was performed by 65.2% of the respondents before touching a patient and 54.5% after touching the patient surroundings while 13.6% performed only when it was visibly soiled. Mobile phones and stethoscopes were disinfected by 13.6 and 30.3% of the respondents after each patient encounter, respectively. Aprons were washed after using them at a stretch for a median duration of 5 days (1-30 days). Forgetfulness, lack of reinforcement, lack of time, inadequate awareness on standard disinfection practices and fear of damaging electronic devices from disinfectants use were reasons for poor compliance. CONCLUSIONS: There is an urgent need to spread awareness and formulate standard guidelines on disinfection practices especially for mobile phones, stethoscopes, and aprons in addition to reinforcing hand hygiene practices.


Subject(s)
COVID-19/prevention & control , Equipment Contamination/prevention & control , Guideline Adherence , Hand Hygiene/standards , Health Personnel/psychology , Cross Infection/prevention & control , Cross-Sectional Studies , Hand Disinfection , Health Facilities , Health Knowledge, Attitudes, Practice , Humans , Intensive Care Units , SARS-CoV-2 , Surveys and Questionnaires
20.
Viruses ; 13(6)2021 05 22.
Article in English | MEDLINE | ID: covidwho-1244142

ABSTRACT

The SARS-CoV-2 pandemic has highlighted the need for protective and effective personal protective equipment (PPE). Research has shown that SARS-CoV-2 can survive on personal protective equipment, such as commonly used surgical masks. Methods are needed to inactivate virus on contaminated material. We show here that embedding viral-disinfecting compounds during the manufacturing of surgical masks inactivates a high dose (up to 1 × 105 pfu) of live, authentic SARS-CoV-2 within minutes.


Subject(s)
Ammonium Compounds/pharmacology , Antiviral Agents/pharmacology , Masks/virology , SARS-CoV-2/drug effects , Virus Inactivation , Equipment Contamination/prevention & control , Humans
SELECTION OF CITATIONS
SEARCH DETAIL